7 Operation
61
7.5.6 Menu:
Input/Output
Menu / Parameters
Description
Input/Output
...Curr.Out 31 / 32
Used to select the “...Curr.Out 31 / 32” submenu with
.
...Dig.Out 41 / 42
Used to select the “...Dig.Out 41 / 42” submenu with
.
...Dig.Out 51 / 52
Used to select the “...Dig.Out 51 / 52” submenu with
.
Input/Output / Curr.Out 31 / 32
Output Value
Used to select the process variable output at the current output.
• Qm [%]: Mass flow
• Qv [%]: Volumetric flow
Curr.Out Mode
Used to select the operating mode for the current output.
• “4-20mA FWD”: Output of flow in upstream direction:
4 mA = no flow
20 mA = maximum flow).
• “4-12-20 mA”: Output of flow in upstream and downstream direction:
4 mA = maximum flow in downstream direction
12 mA = no flow
20 mA = maximum flow in upstream direction
• “4-20mA FWD/REV”: Output of flow in upstream and downstream direction without differentiation of flow direction:
4 mA = no flow
20 mA = maximum flow).
Default setting: 4-20mA FWD/REV
Curr.Out at Alarm
Used to select the state of the current output in the case of a fault.
The output “Low Alarm” or “High Alarm” current is configured in the subsequent menu.
Low Alarm
Used to set the current for the low alarm (3.5 to 3.6 mA).
Default setting: 3.5 mA
High Alarm
Used to set the current for the high alarm (21 to 22.6 mA).
Default setting: 21.8 mA
Curr.Out > 20.5mA
Current output behavior when 20.5 mA are exceeded.
• Hold Last Value: Used to maintain and output the most recent measured value.
• High Alarm: Used to output the current for high alarm.
• Low Alarm: Used to output the current for low alarm.
Default setting: Hold Last Value
Curr.Out < 3.8mA
Current output behavior when below 3.8 mA.
• Hold Last Value: Used to maintain and output the most recent measured value.
• High Alarm: Used to output the current for high alarm.
• Low Alarm: Used to output the current for low alarm.
The parameter is not available if for the “Curr.Out Mode” parameter 4-20mA FWD/REV was selected.
Default setting: Low Alarm
Curr. at EPD Alarm
Current output behavior in the case of an empty meter tube.
• Off: No effect on the current output.
• Q=0 %: The current output is set to 4 mA, which corresponds to “no flow”.
• High Alarm: Used to output the current for high alarm.
• Low Alarm: Used to output the current for low alarm.
Default setting: Off
Содержание 406060
Страница 2: ......
Страница 98: ...13 Appendix 98 13 3 Declarations of conformity 13 3 1 JUMO flowTRANS MAG S10...
Страница 99: ...13 Appendix 99...
Страница 100: ...13 Appendix 100...
Страница 101: ...13 Appendix 101...
Страница 102: ...13 Appendix 102...
Страница 103: ...13 Appendix 103...
Страница 104: ...13 Appendix 104 13 3 2 JUMO flowTRANS MAG H10...
Страница 105: ...13 Appendix 105...
Страница 106: ...13 Appendix 106...
Страница 107: ...13 Appendix 107...
Страница 108: ...13 Appendix 108...
Страница 109: ...13 Appendix 109...
Страница 110: ...13 Appendix 110 13 3 3 JUMO flowTRANS MAG 10...
Страница 111: ...13 Appendix 111...
Страница 112: ...13 Appendix 112...
Страница 113: ...13 Appendix 113...
Страница 115: ......