10 Repair
77
10.3 Replacing the LCD indicator
A
Integral mount design
B
Remote mount design
1
Housing cover
2
Fastening screws for LCD display
3
LCD display
4
Connector
5
Four fastening screws for housing cover
Figure 51: Replacing the LCD display
Change from one to two columns
NOTICE
If the O-ring gasket is seated incorrectly or damaged, this
may have an adverse effect on the housing protection
class.
Follow the instructions in
Opening and closing the terminal
box
on page 22 to open and close the housing safely.
The LCD display can be replaced in the case of a defect.
Component Part
number
LCD display (HMI).
For integral and remote mount design
types
00706388
Follow the steps below to replace the LCD display:
1.
Switch off the power supply.
2.
Unscrew and remove the cover.
3. Loosen the fastening screws for the LCD display
(applicable to integral mount design only).
4.
Take off LCD display.
5.
Detach connector from motherboard.
6.
Insert connector of new LCD display. Take care not to
damage the cable harness.
7. Insert the LCD display and tighten screws as needed.
8.
Attach and screw on cover again.
9.
Switch on the power supply.
Change from two to one column
Содержание 406060
Страница 2: ......
Страница 98: ...13 Appendix 98 13 3 Declarations of conformity 13 3 1 JUMO flowTRANS MAG S10...
Страница 99: ...13 Appendix 99...
Страница 100: ...13 Appendix 100...
Страница 101: ...13 Appendix 101...
Страница 102: ...13 Appendix 102...
Страница 103: ...13 Appendix 103...
Страница 104: ...13 Appendix 104 13 3 2 JUMO flowTRANS MAG H10...
Страница 105: ...13 Appendix 105...
Страница 106: ...13 Appendix 106...
Страница 107: ...13 Appendix 107...
Страница 108: ...13 Appendix 108...
Страница 109: ...13 Appendix 109...
Страница 110: ...13 Appendix 110 13 3 3 JUMO flowTRANS MAG 10...
Страница 111: ...13 Appendix 111...
Страница 112: ...13 Appendix 112...
Страница 113: ...13 Appendix 113...
Страница 115: ......