Electrical connection pin assignments
42
28.08.2019
6.2.5.
Connecting to the JULABO Pressure / Flow Sensor socket
Signal input for the connection of an external pressure sensor or flow-
through sensor. Please note that only the JULABO accessory sensors
will work with this socket.
Julabo Sensor
Pressure / Flow
Pin
Signal
1
24 V
2
5 V
3
4…20 mA
4
GND (0 V)
Pressure sensors for
i
n JULABO program
Order No.
Description
8980771
Pressure sensor M24x1.5 male
Possible incorrect measurement when connecting a JULABO
pressure sensor!
•
If the customer's application is positioned higher or lower than the
Temperature Control System, the height difference
will affect the measurement and cause incorrect results.
Remedy:
Position the
Temperature Control System and the
application at the same height level.
Check the plausibility of the sensor measurement regularly (about every 2-3 days).