Operating and functional elements
18
28.08.2019
4.1.1.
Electrical connections on the front side of the device (upper section)
Remove the ventilation
grid (front panel at W80)
as follows in order to
access the electrical
connections:
•
Grasp the bottom
edge of the ventilation
grid (front panel), lift
slightly (1), and tilt it
forwards (2).
•
Remove the
ventilation grid (front
panel) from the
housing frame.
Included connections:
1 Ethernet connection.
2 USB device (Type B
plug).
3 RS232 interface
4 Alarm socket for ext.
alarm signal.
5 Socket for external
measurement and
control sensor
(Pt100).
6 Connection for
draining residual bath
fluid.
7 Sensor for ambient
temperature
measurement.
Using tubing and a
suitable vessel, open the
connection (6), about
every 2-3 weeks, to
release any built-up
deposits