A80, A80t, W80, W80t Operating Manual
28.08.2019
15
3.6.2.
Properly locating the device, Earthquake anchorage
•
Transport the unit on a
firm, level surface.
Avoid shocks, e.g. floor
grates.
•
For longer distances,
securely strap the unit
to a trolley or cart before
transport.
•
Position the unit at the
desired location and
lock the castors.
•
Install the unit on a
level, noncombustible
surface.
•
Do not install the unit in
the immediate vicinity of
a heat source and do
not place it in direct
sunlight.
•
Keep the ventilation
grids clear (front and
rear)! Maintain
clearance of at least 8
inches (20 cm) from
objects, walls, etc. in
both the front and rear
of the unit.
•
Clean the ventilation
grids (front and rear) at
least once every 2-3
weeks. In dirty
environments, clean the
ventilation grids more
often.
•
When using an external
reservoir, additional
space is needed at the
rear.
Ensure adequate air conditioning at the installation location.
The installation location should be air conditioned to prevent the
exhaust heat of the unit from increasing the ambient temperature too
much (maximum ambient temperature 40 °C [104 °F]). Furthermore,
standard EN 378 requires a minimum room size for each kilogram of
refrigerant in the event of a leak in the refrigeration circuit. Refer to the
type label for the refrigerant volume.
0.49 kg of refrigerant R507 requires 1 m
3
of space.
0.68 kg of refrigerant R23 requires 1 m
3
of space.