11. Maintenance/repair of the unit
Danger of electric shock
Risk of injury from electric power.
•
Always switch off the device and unplug it from its power supply
before performing any cleaning tasks.
Never allow moisture to collect inside the temperature control
system.
•
Service and repair tasks may be performed only by JULABO-
authorized service providers.
Clean the condenser occasionally in order to keep the unit running with
full cooling capacity.
•
Switch off the unit, and unplug it from its power source.
•
Remove the ventilation grid.
•
Vacuum off dirt from the condenser.
•
Replace the ventilation grid.
11.1. Repair service
We recommend speaking with our technical service department before
requesting a service technician or sending a JULABO device in for repair.
•
Completely drain the unit, remove the expansion tank, seal the
connections with nuts and sealing caps.
•
Package the unit carefully and properly.
•
Always include a brief description of the problem.
If you send your JULABO unit back to us, please use the respective
online return form
•
The unit must be standing upright during shipment.
•
Label the packaging properly to avoid unintentional damage.
JULABO will not be liable for any damage caused by improper
packaging.
During the repair process, JULABO will perform any upgrades or
technical changes that are necessary to ensure the reliable operation of
the device.