Electrical connection pin assignments
38
28.08.2019
6.1.2.
Connection for the external Pt100 sensor
The cable's shield must be electrically connected to both the sensor’s
sheath as well as the housing of the connector plug.
External sensor
connection for external
temperature measurement
and control. Only use
shielded sensors.
Check the plausibility of
the external temperature
measurement regularly
(about every 2-3 days), by
placing the Pt100 in a
state where the
temperature is already
known (e.g. in ambient air,
or a container of ice
water).
ext. Pt100
1
4
2
3
Shield
Plug
Look on
soldering side.
Pt100
Pin
Signal
1
I+
2
U+
3
U-
4
I-
6.2.
Options (on the rear side)
Danger: electricity.
During assembly conductive objects may drop into the unit and cause a
short circuit.
Disconnect the unit from the power supply prior to opening.
Installation and maintenance may be performed only by authorized,
qualified personnel.
7 Socket for connection of
optional accessory
analog connections
(Order No. 8900105).
8 Socket for connection of
Pt100 module with
connector (Order No.
8900106).
9 JULABO Pressure /
Flow Sensor socket.
4
1
3
2