A80, A80t, W80, W80t Operating Manual
28.08.2019
49
Temperature setting
Digital
Local / Remote
Control type displayed on TFT User Interface
Remote Control from PC
Displayed on TFT User Interface
Temperature display
Displayed on TFT User Interface
Resolution
°C
0.01
ATC function
(ATC3) internal
(ATC2) extern.
K
± 3
± 10
Temperature control
ICC – Intelligent Cascade Control
Cascade, parameters may be retrieved and modified
Working temperature sensor
Pt 100 1/3 DIN
Safety temperature sensor
Pt 100
8.1.
Cooling water connection, cooling water quantity
Cooling water connections
G3/4 male / barbed fittings for ½“ I.D. tubing
Maximum water pressure
bar [psi]
6 [87]
Min. water pressure at 30 °C
bar [psi]
0.5 [7.25]
Necessary pressure difference
bar [psi]
0.5 [7.25]
Cooling water quantity
l/min
2.0 (at 1.2 kW cooling capacity)
Max. cooling water temperature
°C [°F]
30 [86]
8.2.
Connections
Computer interfaces
RS232,
USB-Device V2.0 /
Ethernet
USB-Host V2.0
SD Card
USB-Stick
SD-card – up to 2 GB
Alarm output
for external alarm signal
External measurement and control sensor
Pt100
Optional Accessories: (Rear side)
Analog module
REG+E-PROG
Programmer input -100 °C - 400 °C
0 - 10 V or 0 - 20 mA or 4 - 20 mA
Temperature recorder outputs
0 - 10 V (0 V = -100 °C, 10 V = 400 °C)
0 - 20 mA (0 mA = -100 °C, 20 mA = 400 °C)
4 - 20 mA (4 mA = -100 °C, 20 mA = 400 °C)
Stand-By-input
for external switch on/off
JULABO Pressure / Flow Sensor
for pressure sensor / flow sensor
Module with Pt100 connection socket
for connection of a further Pt100
www.
GlobalTestSupply
.com
Find Quality Products Online at: