Operator Responsibilities - General Safety Notices
10
28.08.2019
3.3.
Safety instructions for operation
It is essential that you read the operating manuals before initial
operation.
It is important that you follow all safety instructions in order to avoid
personal injury and property damage. These instructions supplement
workplace safety regulations.
•
This device may be connected to grounded (protected earth, PE)
mains power outlets only!
•
These tasks may be performed by properly trained personnel only.
•
The mains plug serves as a reliable way to disconnect the unit from
its power supply for safety reasons and must be readily accessible
at all times.
•
Install the unit on a level, noncombustible surface
•
Adjust the high temperature cut-off before operating the unit.
•
Never operate the unit without bath fluid!
•
Observe expansion of the bath oil as the bath temperature rises.
•
Do not allow water to enter hot bath oil.
•
Do not drain bath fluid while it is hot!
Check the temperature of the bath fluid before draining; this may be
done by briefly switching on the unit.
•
Use tubing suitable for temperature-control purposes.
•
Avoid kinking the external tubing.
•
Secure tubing connections to prevent slipping.
•
Regularly check tubing for material fatigue, such as cracks.
•
Never operate a damaged or leaking unit.
•
Do not attempt to use the unit if the power cable is damaged!
•
Before performing service or repair tasks or moving the unit, switch
the unit off and remove the power plug from the socket.
•
Allow only authorized technicians to perform service and repair
tasks.
•
Always switch off the device and disconnect it from its power supply
before attempting to clean it.
•
Completely drain the unit before moving it.
•
Transport the unit carefully.
•
Do not tilt or lay the unit during transport.
•
Shaking or falls may damage the inside of the unit.
•
Observe all safety labels!
•
Do not remove safety labels!
Hot or cold parts
.
Risk of burns or frostbite.
Use gloves.
Dangerous electrical voltage
Discharging of internal capacitors takes 5 sec. Do not touch plug pins
for 5 sec. after pulling the power plug.
www.
GlobalTestSupply
.com
Find Quality Products Online at: