A80, A80t, W80, W80t Operating Manual
28.08.2019
29
5.7.2.
Filling a closed, external system
(Also refer to the
"User Interface"
operating manual
, 1.951.3041
)
Connect the unit to the
mains power supply and
switch on the unit at the
mains switch.
After the self-test is
completed, the unit will be
in the "OFF" state and will
emit an audible signal.
To mute the audible
alarm, touch the alarm
notice's red box.
The fill level indicator on
the left side of the display
shows that the unit is
empty.
To continue,
press the
button.
You will see a message
instructing how to fill the
unit. Press the button
labeled “Fill unit” and
follow the instructions
displayed on the User
Interface.
Filling the unit:
•
Open the cover on top
of the housing
•
Remove the plug (not
pictured).
•
Slowly pour heat
transfer liquid into the
round opening (Use a
funnel or a container
designed for pouring,
for best results).
The area around the filling
opening is sealed from the
housing, but you should
avoid spilling large
amounts of bath fluid.
Clean up any spilled fluid
with a towel or similar.