A80, A80t, W80, W80t Operating Manual
28.08.2019
41
6.2.4.
Connecting to the REG+E-PROG socket
The programmer input and all of the datalogger outputs are configurable
via the TFT user-interface.
Channel 1
Voltage output, e.g. for datalogger
(V)
Channel 2
Voltage output, e.g. for datalogger
(V)
Channel 3
Current output, e.g. for datalogger
(mA)
E-PROG
Input for external programmer
With the E-PROG - input, you can use either a voltage source or a current
source to change the unit’s setpoint temperature.
REG+E-PROG
SELV input (Safety Extra
Low Voltage)
The programmer input (E-
PROG, pin 4) can be
configured as a voltage or
current input via the TFT
user interface.
Pin
Signal
1 Voltage output*, channel 1
0 ... 10 V
2 Voltage output*, channel 2
0 ... 10 V
3 Gnd for outputs
0 V
4 Programmer input, EPROG
0 ... 10 V or 0 ... 20 mA
5 Current output**, channel 3
0 ... 20 mA or 4 ... 20 mA
6 Gnd for programmer
0 V
* The resistance of any load placed on either of the voltage
outputs must be, at minimum, 10 k
Ω
.
** The resistance of any load placed on the current output must be,
at minimum, 400
Ω
.