A80, A80t, W80, W80t Operating Manual
28.08.2019
45
8.
Technical Data
A80
Order No.
9420801.14
9420801.03
9420801.04
Mains connection (nom. volt.)
Nominal frequency
Tolerance of the nom.voltage
Max. current consumption
Max. heating capacity at 200 V
at 208 V
at 220 V
at 230 V
V
Hz
%
A
kW
kW
kW
kW
208
60
- 10…+ 15
15.0
1.3
1.5
1.76
-
230
50
±
10
16.0
-
-
-
1.8
230
50
±
10
13.0
-
-
-
1.05
Working temperature range
°C [°F]
-80…250 [-112…482]
Temperature stability
K
± 0.01…± 0.05
Cooling capacity
Liquid: Silicon Oil//Ethanol*
°C
kW
200 100 20
0
-10 -20 -30 -40 -60 -80
1.2 1.2 1.2 1.2 1.2 1.1 1.1 1.1 0.65 0.1
Refrigeration compressor
dual
stage
air-cooled
Refrigerant
First stage: R507; Second Stage: R23
Pump capacity
Pressure, max.
Flow rate, max. at 0 bar
bar [psi]
l/min
[gal/min]
1.3 [18.85]
1.7 [24.66]
38
[10.04]
40
[10.57]
Viscosity not to exceed max. 50 mm
2
/s (50 cSt)
Noise level, 1 m distance
dBA
68
Ambient temperature
°C [°F]
5…40 [41…104]
Rel. Humidity
%
< 80 (at <31
°
C) up to < 50 (at 40
°
C)
Process volume, min.
Active heat exchanger vol.
Internal usable expansion vol.
l [gal]
l [gal]
l [gal]
3.9 [1.04]
1.7 [0.45]
5.6 [1.48]
Overall dimensions (WxDxH)
mm
[inch]
430 x (650 + (Stud 35)) x 1258
[16.9 x (25.6 + Stud 1.38)] x 49.5]
Weight, approx.
Kg [lbs]
164 [361.6]
* at 200 °C [392 °F] and 100 °C [212 °F] : measured using Thermal HL45
All data provided in this operating manual assumes use at the rated nominal voltage and rated
nominal frequency and with ambient temperature of 20 °C [68 °F].
Information subject to technical changes without notice.
www.
GlobalTestSupply
.com