Preparations
34
28.08.2019
5.9.
Additional thermal adjustment and electrical connection elements
5.9.1.
Adjustable high temperature cut-off
Two circular potentiometers for setting the high temperature cut-off
are located on the front of the unit. Use a screwdriver to turn these
dials.
On the TFT you will find
the continuous display of
important values and
functions.
Warning
The >TANK< high
temperature cut-off
(complies with DIN
standard 12876-1-2000)
should be set to 15 °C
above the highest
working temperature
setpoint. Make sure that
the fluid is rated for use
at this temperature, first!
The >RESERVOIR< high
temperature cut-off must
be set to at least below
the bath fluid's flash
point.
Adjust the high
temperature cut-off by
slowly turning the dial
with a screwdriver. The
exact value will appear
on the display.
1 Date / time
2 Status: On/Standby
3 Fill level indicator
4 External temperature sensor value
5 Current power output (X% heating, -X% cooling)
6 Selected temperature control mode (internal/external)
7 Current liquid temperature
8 Selected max. pump pressure and current pump pressure
9 Selected high temperature warning setpoint
10 Selected low temperature warning setpoint
11 High temperature cut-off setting (
TANK
)
12 High temperature cut-off setting (
RES
)
Cooling indicator icon - Blinking or continuous
Heating indicator icon - Blinking or continuous