Preparations
30
28.08.2019
Watch the level indicator.
As soon as the minimum
filling level has been
reached the fill level
indicator will turn green.
Switch the unit off and
switch it back on again.
You may now continue
filling up to the desired
level.
Return to standard display
by pressing
.
If there is too much bath
fluid or if the bath fluid
expands due to heating
during operation, a high fluid
level warning is activated.
In this case, use the residual
fluid drain (6) to remove
some bath fluid.
See the next chapter for
draining instructions..
Ticker text:
The early warning system for high fluid level reports a critical fluid
level. Please drain some bath fluid.
5.8.
“Empty the unit” menu
•
Do not drain the bath fluid while it is hot!
Check the temperature of the bath fluid before draining; this may
be done by briefly switching on the unit.
•
Always store and dispose of old bath fluid in an environmentally
friendly way.
Always observe the applicable disposal regulations in your region.