PAGE 77 of 140
DP0029 Rev 02.2
5.9.4. Pivoting Probe Holder Frame Setup - Longitudinal Scanning
5.9.4.1 Longitudinal Scanning
To prepare the pivoting probe holder frame for longitudinal scanning, follow
these steps:
NOTE:
The swivel mount must be in a horizontal position during
longitudinal scanning (see "Swivel Mount" on page 44).
1.
Loosen the pivot wing knobs at the centre of the frame system
(Fig. 177).
Lower the left side of the frame system to align with the tangent of the
scan surface. Tighten the pivot wing knobs.
Fig. 177 - Loosen pivot wing knobs