
PAGE 58 of 140
1.
Ensure the probe holder is in latched, upper position
(Fig. 120)
.
2.
Using the supplied 3 mm hex driver loosen the transverse adjustment
screw
(Fig. 124)
and rotate the yoke about the vertical shaft achieving the
desired angle.
3.
Tighten the transverse adjustment screw
(Fig. 125)
.
To return the transverse adjustment to neutral
(90°)
. The probe holder must
be in the latched, upper position
(Fig. 120)
. Rotate the yoke until the stop post
contacts the base of the probe holder
(Fig. 126)
. Then tighten the transverse
adjustment screw.
5.8.2.4 Probe Holder Longitudinal Adjustment
To adjust the probe holder’s vertical angle for longitudinal scanning, follow
these steps:
1.
Ensure the probe holder is in latched, upper position
(Fig. 120)
.
2.
Using the supplied 3 mm hex driver
(Fig. 37)
, loosen the longitudinal
adjustment screw
(Fig. 127)
.
3.
Rotate the main body of the probe holder until it is at the desired angle
(Fig. 128)
.
4.
Tighten the longitudinal adjustment screw
(Fig. 128)
.
To return the longitudinal adjustment to neutral (90°). Line up the longitudinal
adjustment indicator markers
(Fig. 129)
.
Fig. 127 - Loosen 3 mm screw
Fig. 128 - Rotate to position
Fig. 129 - Line up markers