BOBCAT Hardware User’s Manual
IMPERX
Rev. 2.0.7
6421 Congress Ave.
4/8/2014
Boca Raton, FL 33487
www.imperx.com
+1 (561) 989-0006
268 of 329
1.
Off
– No de-bounce
2.
10.0
us – 10 microseconds de-bounce interval.
3.
50.0
us – 50 microseconds de-bounce interval.
4.
100.0
us – 100 microseconds de-bounce interval (default).
5.
500.0
us – 500 microseconds de-bounce interval.
6.
1.0
ms – 1 milliseconds de-bounce interval.
7.
5.0
ms – 5 milliseconds de-bounce interval.
8.
10.0
ms – 10 milliseconds de-bounce interval.
Acquisition:
Frames per trigger
– sets the number of frames captured after each
trigger in “Standard” mode. The user can select from 1 to 65500 discrete
frames to be captured. If the value is 65501 or bigger, the camera is free
running after the trigger signal.
Pulses per capture
– sets the number of trigger pulses used during a
single acquisition event in “Frame Accumulation” mode. The user can
select from 1 to 65535 discrete pulses.
Exposure Delay:
Sets the delay between the trigger pulse active edge and beginning of the
exposure. The user can set the delay from 0 to 16777215 us.
5.7.4 Pulse Generator
In this window the user can configure the parameters of the Internal Pulse Generator –
Figure 5.13.
Figure 5.13 – Pulse generator window