BOBCAT Hardware User’s Manual
IMPERX
Rev. 2.0.7
6421 Congress Ave.
4/8/2014
Boca Raton, FL 33487
www.imperx.com
+1 (561) 989-0006
233 of 329
4.3.8.16
Tap Balance
This register enables the tap balance. If the “Tap balance once” is to be used, the
register has to be set every time from “00” to “10”.
Address
:
0x0110
Data (1:0)
:
00 – no tap balance
01 – dynamic automatic tap balance
10 – dynamic balance taps once
11 – static automatic tap balance
Data (31:2)
:
N/A
4.3.9 Triggering Workspace Registers
4.3.9.1
Trigger Input Selector
This register selects the triggering source.
Address
:
0x0508
Data (2:0)
:
000 – off – no trigger, free running mode – CL only
(this is not available for GigE cameras)
001 – external – the camera expects the trigger to come from
the external source mapped to the power and I/O
connector.
010 – internal – the camera expects the trigger to come from
the programmable pulse generator.
011 – computer – the camera expects the trigger to come
from the camera link cable.
100 – software trigger – expects a one clock cycles pulse
generated by the computer. The trigger exposure is
internal register controlled. Pulse duration exposure is
not allowed.
101 to 111 – N/A
Data (31:3)
:
N/A
4.3.9.2
Trigger Input Mode – GigE cameras
This register enables or disables the triggering operation for GigE cameras.
Address
:
0x05BC
Data (0)
:
1 – trigger is disabled, free running mode
0 – trigger is enabled – camera is in trigger mode
Data (31:1)
:
N/A
4.3.9.3
Software Trigger Start