Determining System Measurement Uncertainties
In any measurement, certain measurement errors associated with the system add
uncertainty
to the measured results This uncertainty
how accurately a device under test
can
be measured.
Network analysis measurement errors can be separated
two types: raw and residual.
The raw error terms are the errors associated
the uncorrected system that are called
systematic (repeatable), random (non-repeatable), and drift errors. The residual error terms are
the errors that remain after a measurement calibration.
The error correction procedure,
called measurement calibration, measures a set of
calibration devices
known characteristics. It uses the measurement results to effectively
remove systematic errors, using the vector math capabilities of the analyzer. The residual
systematic errors remain after error correction, primarily due to the limitations of how
accurately the electrical characteristics of the calibration devices can be defined and
determined. Also, the random (non-repeatable) and drift errors, cannot be corrected because
they cannot be quantified and measured during the measurement calibration and device
measurement. However, the effects of random errors can be reduced through averaging. The
averaging may then be reduced for device measurement. The residual systematic errors along
with the random and drift errors continue to affect measurements after error correction,
adding an uncertainty to the measurement results Therefore, measurement uncertainty is
defined as the combination of the residual systematic (repeatable), random (non-repeatable),
and drift errors in the measurement system after error correction.
The following measurement uncertainty equations and system error models (flowgraphs) show
the relationship of the systematic, random, and drift errors. These are useful for predicting
measurement performance.
Sources of Measurement Errors
The residual (after measurement calibration) systematic errors result from imperfections
in the calibration standards, the connector interface, the interconnecting cables, and the
instrumentation. All measurements are affected by dynamic accuracy and frequency error
effects For reflection measurements, the associated residual errors are effective directivity,
effective source match, and effective reflection tracking. For transmission measurements,
the additional residual errors are effective crosstalk, effective load match, and effective
transmission tracking.
The listing below shows the abbreviations used for residual systematic errors that are in the
error models and uncertainty equations.
n
Efd, Erd = effective directivity
n
Efs, Ers effective source match
n
Efr, Err effective reflection tracking
Determining System Measurement Uncertainties
Содержание HP 8753E
Страница 61: ......
Страница 138: ...Figure 2 62 aw000024 Gating Effects in a Frequency Domain Example Measurement Making Measurements 2 87 ...
Страница 150: ......
Страница 174: ...Figure 3 23 Example Swept Power Conversion Compression Measurement 3 32 Making Mixer Measurements ...
Страница 187: ......
Страница 197: ...4 l 8 Printing Plotting and Saving Measurement Results ...
Страница 245: ......
Страница 249: ......
Страница 256: ......
Страница 277: ...4 lb enter the number of sweeps press Optimizing Measurement Results 5 55 ...
Страница 399: ......
Страница 475: ......
Страница 476: ...pb611ld 8 2 Menu Maps ...
Страница 477: ......
Страница 478: ...84 Menu Maps ...
Страница 480: ......
Страница 487: ......
Страница 488: ...8 14 MenuMaps ...
Страница 489: ...I r RETURN pg657e Menu Maps 8 15 ...
Страница 565: ... Ittble 9 2 Softkey Locations Softkey Front Panel Access Key B 76 Key Definitions ...
Страница 566: ... Ihble 9 2 Softkey Locations continued Key Definitions 8 77 ...
Страница 567: ...9 79 Key Definitions ...
Страница 568: ...Key Definitions 8 78 ...
Страница 569: ... Ihble 9 2 Softkey Locations continued Softkey A i _ 8 80 Key Definitions ...
Страница 570: ...able 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 9 81 ...
Страница 571: ...Ylhble 9 2 Softkey Locations continued Front Panel Access Key iii3 SJ G J ZiJ GiJ Ical m 9 82 Key Definitions ...
Страница 572: ...lhble 9 2 Softkey Lacations continued Softkey Front Panel Access Key Key Definitions 9 83 ...
Страница 573: ... I ble 9 2 Softkey Locations continued Softkey Front Panel Access K e y 84 I ey Definitions ...
Страница 574: ... lhble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions g 86 ...
Страница 575: ... lhble 9 2 Softkey Locations continued Softkey FFont Panel Access Key 9 86 Kay Definitions ...
Страница 576: ... ble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 9 87 ...
Страница 577: ... lhble 9 2 Softkey Locations continued Softkey Front Fanel Access Key B 88 Key Definitions ...
Страница 578: ...l hble 9 2 Softkey Locations continued Softkey Key Definitiins 8 W ...
Страница 579: ... lhble 9 2 Softkey Locations continued Softkey Front Rmel Access Key WI Key Definitions ...
Страница 580: ... Ihble 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions Ml ...
Страница 581: ... able 9 2 Softkey Locations continued Key Definitions ...
Страница 582: ...able 9 2 Softkey Locations continued Softkey Front Panel Access Key Key Definitions 943 ...
Страница 583: ...7hble 9 2 Softkey Locations continued I Softkey Front Panel Access Key 944 Key Definitions ...
Страница 584: ... Ihble 9 2 Softkey Locations continued Softkey Key Definitions 9 96 ...
Страница 671: ......
Страница 672: ......
Страница 673: ......