Temperature Control
Oven temperature calibration
54
INITIAL
TIME
RATE
OVEN
FINAL
TIME
STATUS
RUN
NOT
READY
ACTUAL
SETPOINT
PASSED
SELF
TEST
Message Display, Power Failure and Recovery
Figure 3-8
Heated zones return to their respective setpoint values, after which the
oven returns to its setpoint value.
C
If
OVEN TEMP
is displayed after recovery from a power failure, and if
the oven was
ON
before the power failure, the oven display shows the
actual oven temperature value, and cycles between showing the
setpoint value and
OFF
until other zones achieve their respective
setpoint temperatures.
C
The oven can be switched
ON
through the keyboard at any time,
without waiting for heated zones to first come to correct temperature.
C
An analytical or column compensation run in progress at the time of a
power failure is aborted; similarly, a keyboard entry in progress is
aborted.
Oven temperature calibration
To maximize precision with respect to retention time information,
particularly if retention times are to be compared between
chromatographs, it may be necessary to calibrate oven temperature in the
range of interest using an independent temperature•measuringdevice.
With the factory•setcalibration difference value of 0 (zero), displayed
oven temperature is accurate to within 1% of the actual temperature,
expressed in
^
K (Kelvin).
Содержание 5890 Series II
Страница 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Страница 8: ...This page intentionally left blank ...
Страница 9: ...1 Columns and Fittings ...
Страница 31: ...2 Keyboard and Displays ...
Страница 42: ...This page intentionally left blank ...
Страница 43: ...3 Temperature Control ...
Страница 57: ...4 Electronic Flow Sensing ...
Страница 66: ...This page intentionally left blank ...
Страница 67: ...5 Signal Output ...
Страница 98: ...This page intentionally left blank ...
Страница 99: ...6 Inlet Systems ...
Страница 123: ...7 Detector Systems ...
Страница 155: ...8 Preventive Maintenance ...
Страница 201: ...9 Chromatographic Troubleshooting ...
Страница 222: ...This page intentionally left blank ...
Страница 223: ...10 Test Sample Chromatograms ...
Страница 240: ...This page intentionally left blank ...