Temperature Control
Valid setpoint ranges
46
Valid setpoint ranges
Table 3•1lists valid setpoint ranges for the 13 keys controlling oven and
heated zone temperatures.
NOTE: TOTAL run time will not exceed 650.00 minutes regardless of values entere
,
, and
.
INIT TIME
RATE
FINAL TIME
-80 to 450
1
^
C
Oven Control
-80 to 450
1
^
C
Oven Control
0 to 650.00
0.01 minute
Oven Control
0 to 70
0.1 /minute
Oven Control
-80 to 450
1
^
C
Oven Control
0 to 650.00
0.01 minute
Oven Control
70 to 450
1
^
C
Oven Control
0 to 200.00
0.01 minute
Oven Control
0 to 400
1
^
C
Zone Control
0 to 400
1
^
C
Zone Control
0 to 400*
1
^
C
Zone Control
0 to 400*
1
^
C
Zone Control
0 to 400
1
^
C
Zone Control
Valid
Setpoint Range
In
Increments Of
Function
Key
OVEN TEMP
INIT TIME
INIT TEMP
FINAL TEMP
INJ A TEMP
INJ B TEMP
FINAL TIME
RATE
DET A TEMP
DET B TEMP
EQUIB TIME
OVEN MAX
AUX TEMP
*The valid setpoint range for a Flame Ionization Detector is 0 to 450
^
C.
Table 3-1. Valid Setpoint Ranges For Temperature Control Keys
Содержание 5890 Series II
Страница 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Страница 8: ...This page intentionally left blank ...
Страница 9: ...1 Columns and Fittings ...
Страница 31: ...2 Keyboard and Displays ...
Страница 42: ...This page intentionally left blank ...
Страница 43: ...3 Temperature Control ...
Страница 57: ...4 Electronic Flow Sensing ...
Страница 66: ...This page intentionally left blank ...
Страница 67: ...5 Signal Output ...
Страница 98: ...This page intentionally left blank ...
Страница 99: ...6 Inlet Systems ...
Страница 123: ...7 Detector Systems ...
Страница 155: ...8 Preventive Maintenance ...
Страница 201: ...9 Chromatographic Troubleshooting ...
Страница 222: ...This page intentionally left blank ...
Страница 223: ...10 Test Sample Chromatograms ...
Страница 240: ...This page intentionally left blank ...