Detector Systems
Nitrogen-phosphorus detector (NPD)
132
Performance considerations
Contamination
Very little contamination can create serious NPD problems. Common
sources include:
C
Columns and/or glass wool treated with H
2
PO
3
(phosphoric acid)
C
Phosphate•containingdetergents
C
Cyano•substitutedsilicone columns (XE•60,OV•225,etc.)
C
Other nitrogen•containingliquid phases
C
Any liquid phase deactivated for analysis of basic compounds
C
Fingerprints
C
Leak•detectionfluids
C
Laboratory air
Contamination in the NPD can be either a positive type or a negative
type of contamination. A positive type of contamination is one that gives
a more positive offset than would normally result with a clean system.
When this happens there is a tendency to try to operate the detector with
too little power to the active element, with the final result that the
detector does not appear as sensitive as desirable (because the
temperature of the active element is actually less than normal).
If the contamination is a negative type of contamination, then there is a
tendency to
quench
the reactions that give rise to the specific response
desired. This
quenching
can be only partial, resulting in less sensitivity
than expected, or at very high contamination levels, a complete
quenching of all signals from the detector. This latter type of quenching is
(usually) only seen on sample peaks where the apex of the peak actually
goes back down toward the baseline.
Содержание 5890 Series II
Страница 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Страница 8: ...This page intentionally left blank ...
Страница 9: ...1 Columns and Fittings ...
Страница 31: ...2 Keyboard and Displays ...
Страница 42: ...This page intentionally left blank ...
Страница 43: ...3 Temperature Control ...
Страница 57: ...4 Electronic Flow Sensing ...
Страница 66: ...This page intentionally left blank ...
Страница 67: ...5 Signal Output ...
Страница 98: ...This page intentionally left blank ...
Страница 99: ...6 Inlet Systems ...
Страница 123: ...7 Detector Systems ...
Страница 155: ...8 Preventive Maintenance ...
Страница 201: ...9 Chromatographic Troubleshooting ...
Страница 222: ...This page intentionally left blank ...
Страница 223: ...10 Test Sample Chromatograms ...
Страница 240: ...This page intentionally left blank ...