Columns and Fittings
Liners/adapters and inserts, general
18
1/8-inch Metal
1/4-inch Metal
1/4-inch Glass
Recommended
1/8-inch
1/4-inch
1/4-inch swage-
Column Fittings
swage-type nut
swage-type nut
type nut and
and ferrules
3
and ferrules
3
graphic ferrule or
silicone O-ring(s)
Packed Column
19243-80510
1
19243-80520
1
None
Inlet Liners
or
or
19243-80530
1
19243-80540
1
(requires glass insert)
(requires glass insert)
FID/NPD
2
19231-80521
1
19231-80530
1
None
Liners/Adapters
TCD Liners/Adapters
None
19302-80020
1
19302-80020
1
(may require
altering the
column)
ECD Liner/Adapters
19301-80530
1
None
None
Packed Columns
1
Use 1/4-inch swage-type nut and Vespel or graphite ferrule to install liner/adapter.
2
See Chapter 8 for details regarding jet exchange (if necessary).
3
See information later in this chapter regarding proper installation of swage-type fittings on packed
metal columns.
Table 1-2.
Hardware and Recommended Fittings for Packed Column Installation
Содержание 5890 Series II
Страница 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Страница 8: ...This page intentionally left blank ...
Страница 9: ...1 Columns and Fittings ...
Страница 31: ...2 Keyboard and Displays ...
Страница 42: ...This page intentionally left blank ...
Страница 43: ...3 Temperature Control ...
Страница 57: ...4 Electronic Flow Sensing ...
Страница 66: ...This page intentionally left blank ...
Страница 67: ...5 Signal Output ...
Страница 98: ...This page intentionally left blank ...
Страница 99: ...6 Inlet Systems ...
Страница 123: ...7 Detector Systems ...
Страница 155: ...8 Preventive Maintenance ...
Страница 201: ...9 Chromatographic Troubleshooting ...
Страница 222: ...This page intentionally left blank ...
Страница 223: ...10 Test Sample Chromatograms ...
Страница 240: ...This page intentionally left blank ...