Inlet Systems
Split/splitless capillary inlet
119
10
20
30
40
50
60
Area
Counts
+ 1.2% Deviation
+ 1.2% Deviation
~
20 ppm n-C
14
(Cold Trapped)
~
10 ppm n-C
11
(Solvent Effect)
Purge Activation Time, Sec
Solvent: Isooctane
Column: 16.5 m x 0.25 mm SE-54
80
^
C (0.5 min)
170
^
C @ 15
^
/min
Sample Size: 1.3 ml (Manual Injection)
Injection Rate: 1 ml/sec
Purge Flows: 5 ml/min, Septum: 60 ml/min, Inlet
Effect of Inlet Purge Activation Time on Area Counts
Figure 6-10
Noting Figure 6•10,waiting too long does not increase component peak
areas, but does increase interference by the solvent tail. Purging too early
risks venting light components, not allowing sufficient time for heavier
components to enter the column, and/or not having sufficient solvent
enter the column to ensure good reconcentration.
A recommended procedure is to analyze a known standard using
conditions identical to those to be used for later sample analyses. Ideally,
the standard should be representative (in both components chosen and
their respective concentrations) of unknown samples to be analyzed.
Perform a series of analyses with increasingly delayed insert purge times:
the optimal time is where maximum area counts are obtained with least
interference by the solvent.
Содержание 5890 Series II
Страница 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Страница 8: ...This page intentionally left blank ...
Страница 9: ...1 Columns and Fittings ...
Страница 31: ...2 Keyboard and Displays ...
Страница 42: ...This page intentionally left blank ...
Страница 43: ...3 Temperature Control ...
Страница 57: ...4 Electronic Flow Sensing ...
Страница 66: ...This page intentionally left blank ...
Страница 67: ...5 Signal Output ...
Страница 98: ...This page intentionally left blank ...
Страница 99: ...6 Inlet Systems ...
Страница 123: ...7 Detector Systems ...
Страница 155: ...8 Preventive Maintenance ...
Страница 201: ...9 Chromatographic Troubleshooting ...
Страница 222: ...This page intentionally left blank ...
Страница 223: ...10 Test Sample Chromatograms ...
Страница 240: ...This page intentionally left blank ...