Detector Systems
FID and NPD jets
125
Supply pressure for capillary makeup gas should be set to about 276 kPa
(40 psi).
FID and NPD jets
Depending upon the column type to be used, and/or analyses to be
performed, exchanging the jet in an FID or NPD may be necessary. Table
7•1lists available jets.
Note:
If switching from packed column operation to capillary operation,
the jet for capillary use must be installed
prior
to column installation.
18789-80070
0.030
Packed Column Only
(FID only: Simulated Distillation,
TCD-to-FIS series operation.)
18710-20119
0.018
Packed Column (Standard, FID and NPD)
19244-80560
0.011
Capillary Column (FID and NPD)
(FID: high sensitivity, packed column)
* Measured at the jet tip.
Use
Jet Tip ID (inch)*
Part No.
Table 7-1. Available FID / NPD Jets
For an FID, the 0.011•inchjet maximizes detector sensitivity and
must
be used with capillary columns; if used with packed columns, flame•out
and clogging problems may occur.
For an NPD, exchanging jets is less critical to detector sensitivity: the
0.011•or 0.018•inchjet may be used with either packed or capillary
columns. The 0.011•inchjet
must
be used with capillary columns.
In packed column applications, the 0.011•inchjet may be tried first. The
0.018•inchjet may then be exchanged if problems are encountered. See
Chapter 8
,
Preventive Maintenance, for specific details regarding jet
exchange.
Содержание 5890 Series II
Страница 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Страница 8: ...This page intentionally left blank ...
Страница 9: ...1 Columns and Fittings ...
Страница 31: ...2 Keyboard and Displays ...
Страница 42: ...This page intentionally left blank ...
Страница 43: ...3 Temperature Control ...
Страница 57: ...4 Electronic Flow Sensing ...
Страница 66: ...This page intentionally left blank ...
Страница 67: ...5 Signal Output ...
Страница 98: ...This page intentionally left blank ...
Страница 99: ...6 Inlet Systems ...
Страница 123: ...7 Detector Systems ...
Страница 155: ...8 Preventive Maintenance ...
Страница 201: ...9 Chromatographic Troubleshooting ...
Страница 222: ...This page intentionally left blank ...
Страница 223: ...10 Test Sample Chromatograms ...
Страница 240: ...This page intentionally left blank ...