Preventive Maintenance
Packed column inlet
162
Packed Column Inlet, Leak-Checking the Septum
Figure 8-3
Cleaning
Turn off the heated zone for the inlet and allow it to cool.
Remove the septum retainer nut and septum; remove also the column
and inlet liner. Using a suitable light source, illuminate the inside of the
inlet from inside the oven while looking through the inlet from the top. If
there is evidence of contamination or deposits, the inlet should be
cleaned.
Use a swab and an appropriate solvent to scrub interior walls until
deposits are removed. A suitable wire may be used to dislodge solid
particles carefully. Dry thoroughly with compressed, filtered, dry air or
N
1
before reassembling.
To avoid possible eye injury, wear eye protection when using
compressed gas.
WARNING
Содержание 5890 Series II
Страница 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Страница 8: ...This page intentionally left blank ...
Страница 9: ...1 Columns and Fittings ...
Страница 31: ...2 Keyboard and Displays ...
Страница 42: ...This page intentionally left blank ...
Страница 43: ...3 Temperature Control ...
Страница 57: ...4 Electronic Flow Sensing ...
Страница 66: ...This page intentionally left blank ...
Страница 67: ...5 Signal Output ...
Страница 98: ...This page intentionally left blank ...
Страница 99: ...6 Inlet Systems ...
Страница 123: ...7 Detector Systems ...
Страница 155: ...8 Preventive Maintenance ...
Страница 201: ...9 Chromatographic Troubleshooting ...
Страница 222: ...This page intentionally left blank ...
Страница 223: ...10 Test Sample Chromatograms ...
Страница 240: ...This page intentionally left blank ...