Chromatographic Troubleshooting
Pressure check
217
Pressure check
The pressure•checkmethod will indicate, but sometimes not isolate, a
leak in the flow path. Since this method does not necessarily isolate the
leak, one of the leak•checkmethods may be needed to locate the leak
specifically.
Note that each valve in a system has two flow paths,
on
and
off
. A leak
sometimes occurs in only one of these two positions. Check both.
1. Disconnect the detector from the valve system.
2. Cap the valve system at its outlet and pressurize to 689 kPa (100 psi).
Allow 2 to 5 minutes for pressure to equilibrate. (If a flow sensor
exists, it should read zero flow.)
3. Turn the knob on the regulator counterclockwise until it turns freely.
The regulator is now turned off and the gauge is indicating pressure
within the valve system.
4. Commonly, the pressure will drop quickly for approximately 30-60
seconds; then stabilize. After this initial drop, the gauge should not
show more than a 6.89 to 13.78 kPa (1 to 2 psi) drop during a
10•minuteperiod.
5. If no leak is indicated, actuate all valves and repeat steps 2 through 4.
6. If a leak does show up, try to pinpoint the source with a soap•bubble
technique. Do not assume the leak must exist only at a valve. Often
plumbing connections such as unions or bulkhead fittings are at fault.
7. If the leak cannot be found easily, divide the system in half and repeat
the pressure check. Continue dividing by halves, and pressure check
until the leak is isolated.
Содержание 5890 Series II
Страница 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Страница 8: ...This page intentionally left blank ...
Страница 9: ...1 Columns and Fittings ...
Страница 31: ...2 Keyboard and Displays ...
Страница 42: ...This page intentionally left blank ...
Страница 43: ...3 Temperature Control ...
Страница 57: ...4 Electronic Flow Sensing ...
Страница 66: ...This page intentionally left blank ...
Страница 67: ...5 Signal Output ...
Страница 98: ...This page intentionally left blank ...
Страница 99: ...6 Inlet Systems ...
Страница 123: ...7 Detector Systems ...
Страница 155: ...8 Preventive Maintenance ...
Страница 201: ...9 Chromatographic Troubleshooting ...
Страница 222: ...This page intentionally left blank ...
Страница 223: ...10 Test Sample Chromatograms ...
Страница 240: ...This page intentionally left blank ...