Preventive Maintenance
Split/splitless capillary inlets
164
1. Loosen and remove the septum retainer nut. Remove and discard the
old septum, found either in the top of the inlet or inside the septum
retainer nut.
Capillary Inlet
Septum Replacement, Split/Splitless and Split-Only Capillary Inlet
Figure 8-4
2. The new septum is placed in the top of the inlet base. Make sure that
sealing surfaces at the top of the inlet and inside the retainer nut are
clean (no particulate matter).
3. Replace the septum retainer nut, tightening it to finger•tightness(as
tight as possible by hand without using any kind of tool). Do not
overtighten; the retainer nut is spring•loadedto provide sufficient
pressure for sealing.
Insert care
Cleaning inlet inserts is discussed under Liner and/or insert care, later
in this chapter.
Содержание 5890 Series II
Страница 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Страница 8: ...This page intentionally left blank ...
Страница 9: ...1 Columns and Fittings ...
Страница 31: ...2 Keyboard and Displays ...
Страница 42: ...This page intentionally left blank ...
Страница 43: ...3 Temperature Control ...
Страница 57: ...4 Electronic Flow Sensing ...
Страница 66: ...This page intentionally left blank ...
Страница 67: ...5 Signal Output ...
Страница 98: ...This page intentionally left blank ...
Страница 99: ...6 Inlet Systems ...
Страница 123: ...7 Detector Systems ...
Страница 155: ...8 Preventive Maintenance ...
Страница 201: ...9 Chromatographic Troubleshooting ...
Страница 222: ...This page intentionally left blank ...
Страница 223: ...10 Test Sample Chromatograms ...
Страница 240: ...This page intentionally left blank ...