Electronic Flow Sensing
Designating gas type
59
Designating gas type
To scale the displayed flow rate value properly, one of four commonly used
gases must be designated. The appropriate gas type is selected according
to Table 4•1:
1
He (Helium)
TCD
2
N
2
(Nitrogen)
General
3
H
2
(Hydrogen)
Capillary
4
Ar/CH
4
(Methane in Argon)
ECD
Number
Gas Type
Preferred Use
Table 4-1. Defining Type of Gas to be Monitored
To select one of these gases for a particular flow channel, press:
FLOW
(
B
) to display
FLOW A
(or
FLOW B
).
1
,
2
,
3
, or
4
is then pressed, followed by
ENTER
. Upon pressing
ENTER
, the current flow rate is displayed, scaled appropriately for the
chosen gas type.
If a gas
other
than one of the above
standard
four is used, select He,
N
1
, H
1
, or Ar/CH
3
according to which one is
closest
in thermal
conductivity to the gas being used. Under
no
circumstances should any
corrosive gas be passed through the EFS.
The maximum usable range for H
1
is 100 ml/minute. At higher flow rates
( >100 ml/min), where a gas other than He, N
1
, or Ar/CH
3
is being used,
or to ensure maximum accuracy in displayed flow rate, calibration of the
EFS may be necessary.
Содержание 5890 Series II
Страница 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Страница 8: ...This page intentionally left blank ...
Страница 9: ...1 Columns and Fittings ...
Страница 31: ...2 Keyboard and Displays ...
Страница 42: ...This page intentionally left blank ...
Страница 43: ...3 Temperature Control ...
Страница 57: ...4 Electronic Flow Sensing ...
Страница 66: ...This page intentionally left blank ...
Страница 67: ...5 Signal Output ...
Страница 98: ...This page intentionally left blank ...
Страница 99: ...6 Inlet Systems ...
Страница 123: ...7 Detector Systems ...
Страница 155: ...8 Preventive Maintenance ...
Страница 201: ...9 Chromatographic Troubleshooting ...
Страница 222: ...This page intentionally left blank ...
Страница 223: ...10 Test Sample Chromatograms ...
Страница 240: ...This page intentionally left blank ...