Signal Output
INET configuration
89
Figure 5•11 shows resulting displays.
INET-HP 5890
Signal Definition
ACTUAL
SETPOINT
GLOBAL
ADDR:
8,31
ACTUAL
SETPOINT
SIG
2
OFF
ACTUAL
SETPOINT
SIG
1
ON
FULL
RANGE
ACTUAL
SETPOINT
SIG
1
ON
RANGED
INET Signal Definition Displays
Figure 5-11
From the displays, the following may be noted:
C
HP 5890 signal channels are designated
SIG 1
or
SIG 2
.
C
ON
indicates the given signal channel is considered active by the
controller; data from this signal channel is transmitted to other
devices on the INET loop.
Similarly,
OFF
indicates the channel is considered inactive; no data
from this signal channel is transmitted to other devices on the loop.
It is important to note that
ON
or
OFF
in this context are strictly
INET definitions, defined at the system controller. They do not, for
example, bear any relationship to whether or not a given detector
assigned to the signal channel is turned on or off.
Содержание 5890 Series II
Страница 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Страница 8: ...This page intentionally left blank ...
Страница 9: ...1 Columns and Fittings ...
Страница 31: ...2 Keyboard and Displays ...
Страница 42: ...This page intentionally left blank ...
Страница 43: ...3 Temperature Control ...
Страница 57: ...4 Electronic Flow Sensing ...
Страница 66: ...This page intentionally left blank ...
Страница 67: ...5 Signal Output ...
Страница 98: ...This page intentionally left blank ...
Страница 99: ...6 Inlet Systems ...
Страница 123: ...7 Detector Systems ...
Страница 155: ...8 Preventive Maintenance ...
Страница 201: ...9 Chromatographic Troubleshooting ...
Страница 222: ...This page intentionally left blank ...
Страница 223: ...10 Test Sample Chromatograms ...
Страница 240: ...This page intentionally left blank ...