Signal Output
HP-IL loopback test
91
ACTUAL
SETPOINT
PASSED
SELF
TEST
ACTUAL
SETPOINT
FAILED
SELF
TEST
ACTUAL
SETPOINT
HPIL
LOOPBACK
TEST
HPIL LOOPBACK TEST Displays
Figure 5-12
The message
PASSED SELF TEST
indicates INET, at least with respect
to the HP 5890, is performing satisfactorily. If
FAILED SELF-TEST
is
displayed, a bad cable may be indicated; install a different INET cable
and repeat the test. If
FAILED SELF-TEST
is displayed again for a second
cable, electronic problems within the HP 5890 are indicated.
Note:
The
loopback test
may be used to check for continuity in an
INET cable; an open cable causes test failure. Verify that the cable is at
fault (rather than the HP 5890) by testing a second cable.
If an intermittent cable problem is suspected, the test may be repeated as
necessary while flexing the cable (particularly at its plugs). An ohmmeter
should also be used to test for problems; it is a reliable method for testing
continuity.
Содержание 5890 Series II
Страница 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Страница 8: ...This page intentionally left blank ...
Страница 9: ...1 Columns and Fittings ...
Страница 31: ...2 Keyboard and Displays ...
Страница 42: ...This page intentionally left blank ...
Страница 43: ...3 Temperature Control ...
Страница 57: ...4 Electronic Flow Sensing ...
Страница 66: ...This page intentionally left blank ...
Страница 67: ...5 Signal Output ...
Страница 98: ...This page intentionally left blank ...
Страница 99: ...6 Inlet Systems ...
Страница 123: ...7 Detector Systems ...
Страница 155: ...8 Preventive Maintenance ...
Страница 201: ...9 Chromatographic Troubleshooting ...
Страница 222: ...This page intentionally left blank ...
Страница 223: ...10 Test Sample Chromatograms ...
Страница 240: ...This page intentionally left blank ...