Detector Systems
Thermal conductivity detector (TCD)
146
Optimizing performance
The following sections aid in choosing operating parameters to obtain
optimal TCD performance.
Temperature
TCD sensitivity increases as the temperature difference between the
detector filament (automatically set) and the surrounding detector body
(chosen detector zone temperature) increases.
To minimize changes in detector sensitivity with different operating
temperatures, filament temperature is maintained at a relatively
constant difference above detector operating temperature. Filament
resistance (proportional to temperature) is monitored, and power is
adjusted as needed to maintain the filament at constant temperature.
500
400
300
200
100
0
100
200
30
0
400
Detector Temperature (Body
Temp)
^
C
Block Temperature
50
^
C
T
135
^
C
0
Filament Temperature
Filament
Temp
(
^
C)
Filament Temperature versus Detector Block Temperature
Figure 7-9
Содержание 5890 Series II
Страница 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Страница 8: ...This page intentionally left blank ...
Страница 9: ...1 Columns and Fittings ...
Страница 31: ...2 Keyboard and Displays ...
Страница 42: ...This page intentionally left blank ...
Страница 43: ...3 Temperature Control ...
Страница 57: ...4 Electronic Flow Sensing ...
Страница 66: ...This page intentionally left blank ...
Страница 67: ...5 Signal Output ...
Страница 98: ...This page intentionally left blank ...
Страница 99: ...6 Inlet Systems ...
Страница 123: ...7 Detector Systems ...
Страница 155: ...8 Preventive Maintenance ...
Страница 201: ...9 Chromatographic Troubleshooting ...
Страница 222: ...This page intentionally left blank ...
Страница 223: ...10 Test Sample Chromatograms ...
Страница 240: ...This page intentionally left blank ...