Inlet Systems
Split/splitless capillary inlet
118
A recommended procedure is to perform a series of analyses at
increasingly higher inlet temperatures using components representative
of those of interest, and analyzed using the conditions for later sample
analyses. The optimum temperature is where maximum area counts are
obtained,
and
there is no evidence of thermal degradation products.
Solvent purity
Reconcentration traps any solvent impurities at the head of the column,
which may interfere with peaks of interest. A solvent blank run to check
solvent purity is strongly recommended
before
the solvent is used to
dilute samples for analysis.
Inlet insert purge
After injection, sufficient time is allowed for solvent and sample
components to reconcentrate at the head of the column; then solvent
vapor within the inlet insert is vented (purged).
The specific time delay after injection for insert purging to occur depends
upon the components, solvent, and injection volume, flow rate through
the inlet, and internal insert volume (approximately 1 ml); generally, a
time within the range 0.3 to 2 minutes is reasonable.
Содержание 5890 Series II
Страница 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Страница 8: ...This page intentionally left blank ...
Страница 9: ...1 Columns and Fittings ...
Страница 31: ...2 Keyboard and Displays ...
Страница 42: ...This page intentionally left blank ...
Страница 43: ...3 Temperature Control ...
Страница 57: ...4 Electronic Flow Sensing ...
Страница 66: ...This page intentionally left blank ...
Страница 67: ...5 Signal Output ...
Страница 98: ...This page intentionally left blank ...
Страница 99: ...6 Inlet Systems ...
Страница 123: ...7 Detector Systems ...
Страница 155: ...8 Preventive Maintenance ...
Страница 201: ...9 Chromatographic Troubleshooting ...
Страница 222: ...This page intentionally left blank ...
Страница 223: ...10 Test Sample Chromatograms ...
Страница 240: ...This page intentionally left blank ...