Preventive Maintenance
Nitrogen-phosphorus detector (NPD)
186
Reinstallation
1. Reinstall the jet in the detector base (using a 1/4•inchnut driver).
Make sure that the threads are clean and free of burrs that could
cause damage. If there is any binding, the cause should be
determined and corrected before proceeding. If a torque wrench is
available, no more than 1 Nm (newton meter) of torque should be
used to install a new jet.
Caution
Do not overtighten the jet! Overtightening may permanently damage
the jet, the detector base, or both.
2. Cleaning and reassembly of the collector/collector assembly is as
follows:
C
Powdery deposits may be removed from the interior of the collector by
gently blowing out the loose material with low•pressurecompressed air
or nitrogen.
Caution
Do not attempt to clean out the inside of the collector by inserting
objects such as wires or brushes•to do so may damage the active
element.
C
Inspect the collector assembly to make sure that the active element is
centered in the tube and is not shorted to the side and that it is
positioned horizontally.
Содержание 5890 Series II
Страница 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Страница 8: ...This page intentionally left blank ...
Страница 9: ...1 Columns and Fittings ...
Страница 31: ...2 Keyboard and Displays ...
Страница 42: ...This page intentionally left blank ...
Страница 43: ...3 Temperature Control ...
Страница 57: ...4 Electronic Flow Sensing ...
Страница 66: ...This page intentionally left blank ...
Страница 67: ...5 Signal Output ...
Страница 98: ...This page intentionally left blank ...
Страница 99: ...6 Inlet Systems ...
Страница 123: ...7 Detector Systems ...
Страница 155: ...8 Preventive Maintenance ...
Страница 201: ...9 Chromatographic Troubleshooting ...
Страница 222: ...This page intentionally left blank ...
Страница 223: ...10 Test Sample Chromatograms ...
Страница 240: ...This page intentionally left blank ...