Keyboard and Displays
Entering setpoints
36
CLEAR
can be used anytime during an entry, prior to pressing
ENTER
, to
erase the entry in progress. The
*
disappears, and the original setpoint
display is restored.
Rules
regarding keyboard usage are summarized below:
C
An instrument function key, when pressed, is shown in the display
along with its current setpoint value, and actual value for
continuously monitored functions: signal levels, temperatures, flow
rates.
-
A displayed function is also ready for entering a new setpoint
value simply by pressing appropriate keys.
C
An
*
(asterisk) is flashed as the new value is entered, indicating an
entry in progress.
C
ENTER
must terminate a numeric entry. It is also required in
terminating a sequence to assign a particular detector to a given
output signal channel.
In general, any display showing the flashing
*
must be terminated
using
ENTER
.
Upon pressing
ENTER
, the value is verified to ensure it is within the
range permitted for the particular function, and/or that the value is
internally consistent with previously defined setpoint values for other,
related functions.
-
If the value is accepted,
*
disappears from the display indicating
the new value is stored and implemented. For functions whose
values are continuously monitored,
*
is replaced by the actual
value.
-
If the value is out of range, or inconsistent with another setpoint
value, an appropriate message is displayed. Another setpoint
value may be entered immediately without again pressing the
particular instrument function key.
The original setpoint value remains in force until an acceptable entry
is made.
Содержание 5890 Series II
Страница 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Страница 8: ...This page intentionally left blank ...
Страница 9: ...1 Columns and Fittings ...
Страница 31: ...2 Keyboard and Displays ...
Страница 42: ...This page intentionally left blank ...
Страница 43: ...3 Temperature Control ...
Страница 57: ...4 Electronic Flow Sensing ...
Страница 66: ...This page intentionally left blank ...
Страница 67: ...5 Signal Output ...
Страница 98: ...This page intentionally left blank ...
Страница 99: ...6 Inlet Systems ...
Страница 123: ...7 Detector Systems ...
Страница 155: ...8 Preventive Maintenance ...
Страница 201: ...9 Chromatographic Troubleshooting ...
Страница 222: ...This page intentionally left blank ...
Страница 223: ...10 Test Sample Chromatograms ...
Страница 240: ...This page intentionally left blank ...