1-21
General Information
Specifications
Option A3R
OUT and IN ports
(used for transmit)
Type: Electrical: To ITU-T G.703.
Connectors: BNC, 75
Ω
, unbalanced.
(Small Siemens 75
Ω
unbalanced
option available.)
Rate: 155.52 Mb/s. 51.84 Mb/s.
Line code: 155.52 Mb/s: CMI. 51.84
Mb/s: B3ZS.
Output level: 155.52 Mb/s:
±
0.5 V
±
10%. 51.84 Mb/s: Output level is user
configurable.
STM-0 X CON: 530 mV peak nominal.
STM-0 HI: 1.1 V peak nominal.
STM-0 LOW: 350m V peak
nominal.
Error output: B3 error output pulse on
receipt of STM-0 and STM-1 signals.
TTL pulse termination 75
Ω
or 10 k
Ω
.
Simultaneous STM-1e/STM-1e
and STM-1o: When used in
conjunction with the appropriate optical
interfaces, transmit STM-1 electrical
output signal simultaneously with
optical output signal.
Transmitter
Clock timing: Internal: All rates.
Recovered: From SDH input (CMI or
NRZ electrical or optical).
Ext MTS: 64 kb/s conforming to ITU-T
G.703, 2 Mb/s conforming to ITUT-
G.811. BNC, 75
Ω
, unbalanced or
Siemens (3-pin), 120
Ω
, balanced.
(Siemens (3-pin) connector is present
on option A3R. Option 120 replaces
this connector with a Bantam
connector).
Error Generation
Error type
Single
Rate 10
-N
Other
Frame A1A2
*
N in 4 frame words (STM-1 Only)
B1
*
4 - 9
STM-1 Only
B2
*
3 - 9
STM-1 Only
B3
*
4 - 9
Path FEBE
*
4 - 9
VC-3 Path BIP
*
3 - 9
VC-3 Path FEBE
*
3 - 9
V5 BIP-2
*
3 - 9
V5 FEBE
*
4 - 9
Payload Bit
*
3 - 9
Содержание 37717C
Страница 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Страница 5: ...2 ...
Страница 21: ...xviii Contents ...
Страница 102: ...2 13 Installation Connecting to the Network Location of Laser Apertures on Options 130 and 131 ...
Страница 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Страница 125: ...3 16 Performance Tests PDH Transmitter Output UKK USB UKJ USA Figure 3 7 8 Mb s Isolated Negative Pulse ...
Страница 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Страница 375: ...3 266 Performance Tests Performance Test Record ...
Страница 379: ...4 4 Installation Current Previous Terminology ...
Страница 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Страница 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...