B-1
Appendix B - Fitting/Calibrating/Testing New Modules
Appendix B - Fitting, Calibrating & Testing New
Modules
Some of the Performance Tests for the HP 37717C require additional modules to be
present in the instrument. Table B-1 lists each module (and option) under test and
the additional modules which must be present to test it.
To decide if you need to fit any modules, compare Table B-1 with the options listed
on the Rear Panel of the instrument under test.
If additional modules are required, follow the procedure below when fitting,
calibrating and testing. Some modules also need to be calibrated after fitting (see
Table B-1). Calibration procedures for these modules are included here.
Table B-1
Additional Modules Required for Performance Tests
Module Under Test
Option
Under Test
Module(s)
Required
Part Number
Calibration
Required?
Unstrucured PDH Tx/
Rx
UKK,
[USB]
special 808
None
No
Structured PDH Tx
UKJ, [USA]
SPDH Rx
37717-60021
No
Structured PDH Rx
UKJ,
[USA],
UKL, [USC]
SPDH Tx
37717-60020
No
SDH Tx/Rx
US1, [US5],
A1T, [A1U]
A3R [A3S]
None
No
Optical Tx/Rx
UH1, UH2,
URU, USN,
UKT, 130,
131
SDH Tx/Rx
37714-60048
No
PDH/SDH Jitter Tx
UHK, A3K
PDH Tx
37714-60044*
No
PDH Jitter Rx
UHN [US9],
Jitter Tx PDH
Tx/Rx
37717-60010
37714-60044*
Jitter Tx
Содержание 37717C
Страница 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Страница 5: ...2 ...
Страница 21: ...xviii Contents ...
Страница 102: ...2 13 Installation Connecting to the Network Location of Laser Apertures on Options 130 and 131 ...
Страница 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Страница 125: ...3 16 Performance Tests PDH Transmitter Output UKK USB UKJ USA Figure 3 7 8 Mb s Isolated Negative Pulse ...
Страница 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Страница 375: ...3 266 Performance Tests Performance Test Record ...
Страница 379: ...4 4 Installation Current Previous Terminology ...
Страница 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Страница 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...