3-233
Performance Tests
ETSI/ANSI Receiver Monitor Levels (Option 110)
ETSI/ANSI Receiver Monitor Levels (Option 110)
N OT E
An SPDH Tx Module must be fitted for this test - see Appendix B.
Specifications
Description
The signal from the HP 37717C Option 110 PDH Transmitter is applied to the
Option 110 PDH Receiver after attenuating by an amount equal to the selected
Receiver Flat Loss plus the specified Cable Loss. The Flat Loss is obtained by
inserting a number of 50
Ω
, fixed attenuators in series with a 75/50
Ω
Matching Pad
at one end and a 50/75
Ω
Matching Pad at the other end. The loss of the two
Matching Pads is included in the overall attenuation equation. The Cable Loss is
supplied by inserting the correct Cable Simulators for each bit rate in the attenuation
path.
Equipment Required
Bit Rate
Nominal Loss
Equalization at 1/2 Bit Rate
2.048 Mb/s
20, 26, 30dB
3dB
1.544 Mb/s (DS-1)
20, 26, 30dB
3dB
34.368 Mb/s
20, 26dB
6dB
44.736 Mb/s (DS-3)
20, 26dB
6dB
Table3- 10
2M Bit Rate
Cable Simulator #1
34M Bit Rate
Cable Simulator #3
1.5M Bit Rate
Cable Simulator #5
44M Bit Rate
Cable Simulator #6
Содержание 37717C
Страница 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Страница 5: ...2 ...
Страница 21: ...xviii Contents ...
Страница 102: ...2 13 Installation Connecting to the Network Location of Laser Apertures on Options 130 and 131 ...
Страница 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Страница 125: ...3 16 Performance Tests PDH Transmitter Output UKK USB UKJ USA Figure 3 7 8 Mb s Isolated Negative Pulse ...
Страница 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Страница 375: ...3 266 Performance Tests Performance Test Record ...
Страница 379: ...4 4 Installation Current Previous Terminology ...
Страница 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Страница 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...