3-220
Performance Tests
ETSI/ANSI Internal Transmitter Clock Rates (Option 110)
3. Adjust the Frequency Counter ATTEN and Trigger Level to obtain a stable
reading and ensure that the frequency counter reading is between 44735798.7 Hz and
44736201.3 Hz.
4. Select the HP 37717C TRANSMIT page, and set FREQUENCY OFFSET to
[+20PPM].
5. Adjust the Frequency Counter ATTEN and Trigger Level to obtain a stable
reading and ensure that the frequency counter reading is between 44736693.4 Hz and
44737096.0 Hz.
6. Set the FREQUENCY OFFSET to [-20PPM].
7. Adjust the Frequency Counter ATTEN and Trigger Level to obtain a stable
reading and ensure that the frequency counter reading is between44734904.0 Hz and
44735306.6 Hz.