
3-14
Performance Tests
PDH Transmitter Output (UKK, [USB], UKJ, [USA]
Figure 3-4
2 Mb/s Isolated Negative Pulse
22. Ensure that the ratio of +ve and -ve pulse amplitudes is between 0.95 and 1.05.
23. Ensure that the ratio of +ve and -ve pulse widths is between 0.95 and 1.05.
2.048 Mb/s Balanced Output
24. Connect the Balanced to Unbalanced Converter between the HP 37717C 120
Ω
Balanced SIGNAL OUT port and the Oscilloscope.
25. Select TERMINATION [120
Ω
BAL] on the
display.
26. Repeat steps 13 through 23.
8.448 Mb/s Unbalanced Output
27. Select SIGNAL [8 Mb/s] and TERMINATION [75
Ω
UNBAL] on the
display.
28. Connect the Unbalanced 75
Ω
SIGNAL OUT port to the Oscilloscope Input 1.
Set the oscilloscope termination to 1M
Ω
and press
on the
oscilloscope.
29. Select the 8 Mb G.703 mask on the oscilloscope and store it on the display.
30. Repeat steps 13 through 23 to verify the 8 Mb/s unbalanced output.
TRANSMIT
TRANSMIT
AUTOSCALE
Содержание 37717C
Страница 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Страница 5: ...2 ...
Страница 21: ...xviii Contents ...
Страница 102: ...2 13 Installation Connecting to the Network Location of Laser Apertures on Options 130 and 131 ...
Страница 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Страница 125: ...3 16 Performance Tests PDH Transmitter Output UKK USB UKJ USA Figure 3 7 8 Mb s Isolated Negative Pulse ...
Страница 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Страница 375: ...3 266 Performance Tests Performance Test Record ...
Страница 379: ...4 4 Installation Current Previous Terminology ...
Страница 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Страница 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...