3-11
Performance Tests
PDH Transmitter Output (UKK, [USB], UKJ, [USA]
3. Press
and set up the
display as shown opposite.
4. Press
on the
oscilloscope.
5. Adjust the Oscilloscope Timebase and
Delay to position the positive peak pulse
amplitude at mid-pulse-width point in the
centre of the screen.
6. Verify the Pulse in Figure 3-1 meets the following criteria:
Pulse Amplitude
2.133V to 2.607V
Pulse Width
639 ns to 781 ns
Overshoot
≤
0.474V
Undershoot
≤
0.474V
Figure 3-1
704 kb/s Pulse Criteria
7. Adjust the Oscilloscope Delay to position the negative peak pulse amplitude at
mid-pulse-width point in the centre of the screen and verify that the negative pulse
meets the criteria listed in step 6.
704 kb/s Balanced Output
8. Connect the Balanced to Unbalanced Converter between the HP 37717C 120
Ω
Balanced SIGNAL OUT port and the Oscilloscope, leaving the 75
Ω
termination and
"T" connector in place.
TRANSMIT
AUTOSCALE
Содержание 37717C
Страница 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Страница 5: ...2 ...
Страница 21: ...xviii Contents ...
Страница 102: ...2 13 Installation Connecting to the Network Location of Laser Apertures on Options 130 and 131 ...
Страница 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Страница 125: ...3 16 Performance Tests PDH Transmitter Output UKK USB UKJ USA Figure 3 7 8 Mb s Isolated Negative Pulse ...
Страница 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Страница 375: ...3 266 Performance Tests Performance Test Record ...
Страница 379: ...4 4 Installation Current Previous Terminology ...
Страница 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Страница 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...