3-149
Performance Tests
Received Jitter Accuracy (Options UHN, [US9], A1M [A1Q], A1N [A1R],
A1P [A1S], A3L [A3M], A3V [A3W], A3N [A3P])
Received Jitter Accuracy (Options UHN, [US9], A1M
[A1Q], A1N [A1R], A1P [A1S], A3L [A3M], A3V [A3W],
A3N [A3P])
N OT E
PDH Tx and Rx modules must be fitted for this test - see Appendix B.
Specifications
Overall Accuracy: See main specifications in Section 1.
Description
The HP 37717C Jitter Receiver performance is verified using a special internal
instrument calibration routine. This routine is automatic and tests all parameters
except Hit Count and Demodulated Jitter Output.
Hit Count is tested by connecting PDH Signal Out to PDH Signal In with the
transmitted jitter amplitude just exceeding the receiver Hit Threshold and a check
made to ensure the Hit count is within the modulation frequency accuracy limits.
The Demodulated Jitter Output is verified for accuracy by measuring the amplitude
on an oscilloscope.
Equipment Required
Range 1.6
1.0V/UI
±
10%
Range 16
0.1V/UI
±
10%
Spectrum Analyzer
: HP 8568B
75
Ω
/50
Ω
Matching Pad
: HP 11825B
Calibrated Jitter Generation Module
: See Note.
Oscilloscope
: HP 54503A
75
Ω
Termination
: HP 15522-80010
Содержание 37717C
Страница 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Страница 5: ...2 ...
Страница 21: ...xviii Contents ...
Страница 102: ...2 13 Installation Connecting to the Network Location of Laser Apertures on Options 130 and 131 ...
Страница 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Страница 125: ...3 16 Performance Tests PDH Transmitter Output UKK USB UKJ USA Figure 3 7 8 Mb s Isolated Negative Pulse ...
Страница 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Страница 375: ...3 266 Performance Tests Performance Test Record ...
Страница 379: ...4 4 Installation Current Previous Terminology ...
Страница 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Страница 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...