3-103
Performance Tests
STM-1/STM-4 Optical Interface (Option UH2)
6. Verify that the front panel LASER ON led is on, indicating the laser is enabled.
7. Check the HP 8153A Power Meter reading is between -15 dBm and -8 dBm
(Typically -9 dBm).
8. Press
on the HP 37717C and select SIGNAL [STM-4 OPT].
9. Repeat steps 6 and 7.
10. Press
select STORED SETTINGS NUMBER [0] (Default Settings) and
press
.
11. Verify that the front panel LASER ON led is off before continuing.
12. Set the Optical Attenuator to ATTEN 0 dB, WAVELENGTH 1310 nm ; CAL=0.
Connect the Optical Attenuator between the Power Meter and the HP 37717C Optical
Out port (ensure that all connections are tight and that the cable has no twists).
13. Press
on the HP 37717C and select SIGNAL [STM-1 OPT].
14. Adjust the Optical Attenuator to obtain a reading of -26 dBm on the Power Meter.
15. Press
select STORED SETTINGS NUMBER [0] (Default Settings) and
press
.
16. Verify that the front panel LASER ON led is off before continuing.
17. Disconnect the Optical Attenuator Output from the HP 8153A and connect to the
HP 37717C Optical In Port.
18. Press
then
and select TRANSMITTER AND
RECEIVER [COUPLED].
19. Press
and select SIGNAL [STM-1 OPT].
20. Press
, and then
to start a
measurement.
21. After 5 minutes check that NO TROUBLE is displayed on the
display.
22. Press
select STORED SETTINGS NUMBER [0] (Default Settings) and
press
.
23. Verify that the front panel LASER ON led is off before continuing.
24. Repeat steps 12 through 23, selecting SIGNAL [STM-4 OPT] in steps 13 & 19.
25. Disconnect the test equipment.
TRANSMIT
SDH
OTHER
RECALL
TRANSMIT
SDH
OTHER
RECALL
OTHER
SETTINGS CONTROL
TRANSMIT
SDH
RESULTS
TROUBLE SCAN
RUN/STOP
RESULTS
OTHER
RECALL
Содержание 37717C
Страница 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Страница 5: ...2 ...
Страница 21: ...xviii Contents ...
Страница 102: ...2 13 Installation Connecting to the Network Location of Laser Apertures on Options 130 and 131 ...
Страница 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Страница 125: ...3 16 Performance Tests PDH Transmitter Output UKK USB UKJ USA Figure 3 7 8 Mb s Isolated Negative Pulse ...
Страница 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Страница 375: ...3 266 Performance Tests Performance Test Record ...
Страница 379: ...4 4 Installation Current Previous Terminology ...
Страница 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Страница 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...