3-69
Performance Tests
PDH Receiver Monitor Levels (Special Option 808)
Figure 3-32
PDH Receiver Monitor Levels Test Setup
704 kb/s
4. Press
and set up the display
as shown below.Press
until the
Monitor led is lit.
5. Select
TEST TIMING [SINGLE] [5 SECS]
6. Press
and set up the display
as shown below.
7. Set the Synthesizer to: FREQUENCY -
352 KHz sinewave AMPLITUDE - 237.5
mV pk-pk (26dB down) as measured on the
oscilloscope.
8. Press
and verify that no
errors are displayed at the end of the test
period (5 seconds).
RECEIVE
SIGNAL IN
RESULTS
TIMING CONTROL