3-166
Performance Tests
Jitter Receiver STM-1/STM-4 Optical Interface (Options A1P, [A1S], A3N
[A3P])
WA R N I N G
Safety precautions must be observed when handling the HP 37717C Optical
Modules as these generate laser signals which can cause serious injury. The
guidelines below must be followed;
Check the connector configuration of the Fiber Optic Interfaces. If these are
fitted with a connector interface other than FC/PC then remove the existing
connector interface and fit the FC/PC connector interface.
Check for any damage to the HP 37717C Fiber Optic Interface spring loaded
aperture covers and connectors. Do not power up the instrument if in any doubt
about the integrity of these connectors.
Make all connections to the HP 37717C Fiber Optic Interfaces before powering
up the instrument.
Procedure
1. Connect the STM-1/STM-4 Optical Out Port to the HP 8153A Power Meter via
the Optical Attenuator (ensure that all connections are tight and that the cable has no
twists). Set the Optical Attenuator to ATTEN 0 dB, WAVELENGTH 1310 nm ;
CAL=0.
2. Switch on the HP 37717C, check that immediately on power up the LASER ON
led on the Optical Module illuminates for a few seconds and Recall the HP 37717C
DEFAULT SETTINGS as shown on 3-2.
3. Recall the HP 37717C DEFAULT SETTINGS as shown on 3-2.
4. Recall default settings on the HP 8153A:
Press
to select MENU mode on the HP 8153A.
Press
to display RECALL.
Press
, select 0 —
A and press
to recall the default settings
(wavelength = 1300 nm, measurement time = 200 ms, autorange).
5. Press
then
to select the Power Level measurement on the HP
8153A.
6. Press
on the HP 37717C and select SIGNAL [STM-1 OPT].
7. Verify that the Optical Module LASER ON led is on indicating the laser is
enabled.
8. Adjust the Optical Attenuator to obtain a reading of -28 dBm on the Power Meter.
MODE
SYSTEM
EDIT
EXEC
MODE
dBmw
TRANSMIT
SDH
Содержание 37717C
Страница 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Страница 5: ...2 ...
Страница 21: ...xviii Contents ...
Страница 102: ...2 13 Installation Connecting to the Network Location of Laser Apertures on Options 130 and 131 ...
Страница 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Страница 125: ...3 16 Performance Tests PDH Transmitter Output UKK USB UKJ USA Figure 3 7 8 Mb s Isolated Negative Pulse ...
Страница 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Страница 375: ...3 266 Performance Tests Performance Test Record ...
Страница 379: ...4 4 Installation Current Previous Terminology ...
Страница 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Страница 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...