3-152
Performance Tests
Received Jitter Accuracy (Options UHN, [US9], A1M [A1Q], A1N [A1R],
A1P [A1S], A3L [A3M], A3V [A3W], A3N [A3P])
12. Select
and
setup the display as shown below.
13. Select
TEST TIMING [SINGLE] [10 SECS].
14. Set up the HP 37717C
as
shown opposite.
Press
to start the
measurement.
When the measurement is complete check
the displayed Hit Count is between
211,860 and 216,140.
Demodulated Jitter Output
15. Connect the HP 37717C Demodulated Jitter Output to the oscilloscope
terminated in 75
Ω
.
16. Check that the amplitude of the 10.7 kHz displayed waveform is between 435mV
pk_pk and 565mV pk_pk.
17. Disconnect all test equipment.
TRANSMIT
JITTER
RESULTS
TIMING CONTROL
RESULTS
RUN/STOP