3-144
Performance Tests
Transmitted Jitter/Wander Amplitude Accuracy (Option A3K [A3Q])
Transmitted PDH Jitter Accuracy on Range 80
1. Select
and setup the HP
37717C display as shown opposite.
2. Select
and
setup the display as shown opposite.
3. Connect up the equipment as shown in
Figure 3-50.
Figure 3-50
TX Jitter Accuracy Test Setup PDH Range 80
4. Press the [FREQ] key on the Modulation Analyzer and check that a frequency of
1.024 MHz is displayed. If 1.024 MHz is not displayed ENTER 1.024 MHz via the
keypad.
5. Press [FM]; [3kHz] and [Peak+] keys on the Modulation Analyzer and check that
the result displayed is 6.24 kHz
±
480 Hz.
TRANSMIT
TRANSMIT
JITTER