3-185
Performance Tests
ETSI/ANSI Transmitter Output Level and Waveshape (Option UKZ)
DSX-3 Positive Pulse
1. Recall the HP 37717C Default Settings as shown on 3-2.
2. Select the HP 37717C TRANSMIT
page, select PHYSICAL LAYER and set
as follows;
3. Select the HP 37717C TRANSMIT
page, select ATM LAYER and set as
follows;
4. Connect the UKZ Transmit Module 75 ohms Unbalanced Output to the
Oscilloscope via the 75/50 ohm converter.
5. Select CHAN on the oscilloscope and set to 50 ohms input impedance, DC
coupled.
6. Select CHANNEL 1 on the oscilloscope and select PROBE 2.45 (attenuation
level to compensate for the 75/50 ohm converter).
7. If using the HP 54503A oscilloscope, use the following key sequence to select
and store the correct Telecom Mask.
a. DEFINE MEAS key.
b. COMPARE softkey.
c. TEST ON softkey.
d. WFORM SAVE key.
Содержание 37717C
Страница 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Страница 5: ...2 ...
Страница 21: ...xviii Contents ...
Страница 102: ...2 13 Installation Connecting to the Network Location of Laser Apertures on Options 130 and 131 ...
Страница 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Страница 125: ...3 16 Performance Tests PDH Transmitter Output UKK USB UKJ USA Figure 3 7 8 Mb s Isolated Negative Pulse ...
Страница 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Страница 375: ...3 266 Performance Tests Performance Test Record ...
Страница 379: ...4 4 Installation Current Previous Terminology ...
Страница 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Страница 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...