3 Basic Operation
59
6.
Wash the sensor probe 2 or 3 times in deionized water to remove any dirt, then
submerge the sensor probe in the bottle.
7.
Press the up (
) and down (
) keys to set the DO value to 0.00 mg/L or 0.0%.
8.
Check that "Measurement value" has stabilized, then press the ENTER key to start
calibration.
9.
When the message "Cal complete. Press ENT to Span cal." appears, press the
ENTER key to start the span calibration procedure.
10.
Wash the sensor probe 2 or 3 times with deionized water to remove any dirt, then
submerge the sensor probe in the container filled with the span solution.
11.
Press the up (
) and down (
) keys to set the DO value to the saturated dissolved
oxygen value (mg/L) of the water at that temperature or the dissolved oxygen
saturation ratio.
12.
Check that "Measurement value" has stabilized, then press the ENTER key to start
calibration.
13.
Calibration is finished when the message "Cal complete. ENT to manual cal menu."
appears. Press the ENTER key to return to the calibration parameter selection
screen.