4 Maintenance
84
4
Maintenance
Tip
HORIBA Advanced Techno recommends regular manufacturer maintenance checks in order to ensure
a long product life.
4.1
Contact for maintenance
4.2
Routine care
After measurement
1.
Press the control unit’s POWER key for about 1 second to turn the power OFF.
Note
The operation keys are designed to operate using the pad of a finger, sharp objects can tear the
control unit cover damaging the operation keys.
2.
Remove the sensor guard, and clean the sensor with tap water.
3.
Clean the turbidity sensor with the cleaning brush provided.
4.
Remove the two screws securing the COND guard, and the COND guard itself, and
use a test tube brush to gently remove any dirt from the electrical conductivity
electrode.
5.
Wipe off any dirt with a soft cloth. If parts are very dirty, clean them with neutral
detergent, then rinse them. If parts are contaminated by oil, wipe it off with a soft
cloth soaked in alcohol.
6.
Put the COND guard back in place.
Manufacturer:
HORIBA Advanced Techno Co., Ltd.
31, Miyanonishi-cho, Kisshoin Minami-ku, Kyoto 601-8306, Japan
COND guard